- Intestinal Anastomosis by Use of a Memory-shaped Compression Anastomosis Clip (Hand CAC 30): Early Clinical Experience
-
Hak-Youn Lee, Jin-Hee Woo, Si-Young Park, Nam-Wook Kang, Ki-Jae Park, Hong-Jo Choi
-
J Korean Soc Coloproctol. 2012;28(2):83-88. Published online April 30, 2012
-
DOI: https://doi.org/10.3393/jksc.2012.28.2.83
-
-
4,793
View
-
29
Download
-
8
Citations
-
Abstract
PDF
- Purpose
The safety and the efficacy of the compression anastomosis clip (Hand CAC 30) have been demonstrated by animal studies. This study was designed to evaluate the clinical validity of the Hand CAC 30 in enterocolic side-to-side anastomosis after colonic or enteric resections. MethodsA non-randomized prospective data collection was performed for patients undergoing a side-to-side anastomosis using the Hand CAC 30. Eligibility criteria for the use of the Hand CAC 30 were for anastomoses between the colon and the ileum or between two small bowels. The primary short-term endpoint was the rate of anastomotic leakage. Other clinical outcomes, including intra- and postoperative complications, length of operation time and hospital stay, and the clip elimination time were recorded. ResultsA total of 63 patients (male, 36) underwent an enteric or right-sided colonic resection followed by a side-to-side anastomosis using the Hand CAC 30. Laparoscopic surgery was performed in 36 patients, in whom one patient who underwent a laparoscopic right hemicolectomy was converted to an open procedure (1/32, 3.1%). One patient with ascending colon cancer showed postoperative anastomotic leakage and died of co-morbid ischemic heart disease. There were no other surgical mortalities. The exact date of expulsion of the clip could not be recorded because most patients were not aware of clip elimination. No patients manifested clinical symptoms of anastomotic stricture. ConclusionShort-term evaluation of the Hand CAC 30 anastomosis in patients undergoing enterocolic surgery proved it to be a safe and efficacious alternative to the standard hand-sewn or stapling technique.
-
Citations
Citations to this article as recorded by 
- Comparative early histologic healing quality of magnetic versus stapled small bowel anastomosis
Ana Sofia Ore, Ashley Althoff, David R. Kull, Thomas J. Baldwin, Jeffrey L. Van Eps, Evangelos Messaris Surgery.2023; 173(4): 1060. CrossRef - An Automatic Suturing Machine for Intestinal Anastomosis: Advantages Compared With Hand-Suturing Technique
Sanaz Mosafer Khoorjestan, Gholamreza Rouhi Surgical Innovation.2019; 26(2): 209. CrossRef - Biodegradable anastomotic tube prepared from diblock copolymers of mPEG-b-PLA and mPEG-b-PLGA
Xiaodong Zhang, Zuoqian Yu, Feizhao Jiang, Lechi Ye, Jinlei Li, Xingzhao Ye, Chungen Xing Journal of Macromolecular Science, Part A.2019; 56(7): 667. CrossRef - Anastomosis colorrectales por compresión utilizando el dispositivo NiTi
Quintín Héctor González-Contreras, Mónica de Jesús-Mosso, Jesús Alberto Bahena-Aponte, Omar Aldana-Martínez, Karen Pineda-Solís, Sari Narsil Mejia-Arcadia Cirugía y Cirujanos.2016; 84(6): 482. CrossRef - Laparoscopic colonic anastomosis using a degradable stent in a porcine model
Liang Ma, Xiu-Jun Cai, Hai-Hong Wang, Yan-Lan Yu, Di-Yu Huang, Guang-Ju Ge, Hai-Yi Hu, Shi-Cheng Yu World Journal of Gastroenterology.2016; 22(19): 4707. CrossRef - Colorectal anastomosis using a compression device
Quintín Héctor González-Contreras, Mónica de Jesús-Mosso, Jesús Alberto Bahena-Aponte, Omar Aldana-Martínez, Karen Pineda-Solís, Sari Narsil Mejia-Arcadia Cirugía y Cirujanos (English Edition).2016; 84(6): 482. CrossRef - Long term results of the use of compression anastomosis clips (CAC) in gastrointestinal surgery – the first report
Katarzyna Kuśnierz, Paweł Lampe Polish Journal of Surgery.2015;[Epub] CrossRef - A nickel–titanium memory-shape device for gastrojejunostomy: comparison of the compression anastomosis clip and a hand-sewn anastomosis
Katarzyna Kusnierz, Zdzislaw Lekston, Dmytro Zhavoronkov, Slawomir Mrowiec, Pawel Lampe Journal of Surgical Research.2014; 187(1): 94. CrossRef
|